DETAILS
Ultra-lightweight and portable with a 10 hr battery.
Pre-op/post-op vital signs monitor
The accurate, cost-effective Avante Echo vital signs monitor is available in configurations with your choice of SpO2, NIBP and Temperature. The portable monitor has multi-parameter functionality suited for out-patient care and doctor’s offices.
Features
- NOW with Covidien Quick Temp Technology
- 5.7 inch high resolution display for easy reading
- Flexible configurations
- Built-in rechargeable Lithium-ion battery for 10 hours of continuous working
- Optional thermometer for obtaining temperature in approximately 15 seconds
- Real-time parameters measurement display with trend table for easy reviewing
- Bi-directional communications with central station by wired or wireless network
- Suitable for adult, pediatric and neonatal patient
- Optional Nellcor Oximax™
- PR measurement (from SpO2/ NIBP)
- Optional recorder
SPECIFICATIONS
General
- Size: 200.8mm (L) × 241mm (H) × 189 mm (D)
- Weight: 3 kg
- Display: 5.7 inches,LCD
- Color TFT resolution: 640X480
- Power Supply: 100-240 VAC, 50/60HZ
- Pmax=70VA FUSE T 1.6AL
- Battery Type: Lithium-ion
- Voltage: 14.8 V DC Capacitance: 4,400 mAh
- Working period: Color TFT 480min
- Rechargeable period: < 360min
- Recorder: (Optional)
- Record Width: 48 mm
NIBP
- Method: Oscillometric
- Mode: Manual, Auto, Continuous
- Measuring Interval in AUTO Mode: 1/2/3/4/5/10/15/30/60/90/120/240/480 Min
- Continuous: 5min, interval is 5s
- Measuring Type: Systolic Pressure, Diastolic Pressure, Mean Pressure, Measuring Range
- Adult Mode:
- SYS: 40~270mmHg
- DIA: 10~215mmHg
- MAP: 20~235mmHg
- Pediatric Mode:
- SYS: 40~200mmHg
- DIA: 10~150mmHg
- MAP: 20~165mmHg
- Neonatal Mode:
- SYS: 40~135mmHg
- DIA: 10~100mmHg
- MAP: 20~110mmHg
- Cuff Pressure measuring Range: 0~290mmHg
- Pressure Resolution: 1mmHg
- Maximun mean error: 5mmHg
- Maximum Standard deviation: 8mmHg
- Entire Measuring Period 30~45s typical (depend on HR/motion disturbance)
- Dual Overpressure protection:
- Adult: 297 3mmHg
- Pediatric: 240 3mmHg
- Neonatal: 145 3mmHg
- PR:
- Measuring Range: 40~240bpm
- Resolution: 1bpm
- Accuracy: 3bpm or 3% of the maximum
- IEC 60601-2-30
SpO2
- Resolution: 1 %
- Accuracy
- Adult (including Pediatric):
- 2% 70%~100% SpO2
- Undefined 0~70% SpO2
- Neonate:
- 3% 70%~100% SpO2
- Undefined 0~70% SpO2
- Measuring Range: 0 ~ 100 %
- Alarm Range: 0 ~ 100 %
- Pulse Rate:
- Measuring and Alarm Range: 30 ~ 300bpm
- Resolution: 1bpm
- Accuracy: 3bpm
- Data update period: 2s
- ISO 9919