DETAILS
User-friendly touch-screen patient monitor measures ECG, SpO2, NIBP, heart rate, temperature and respiration. Offers optional dual IBP, and EtCO2
The fully-functional Waveline Touch features a 12.1″ color TFT display with a touch-screen interface. It features multi-lead ECG monitoring, ST segment monitoring and 72 hours trend data analysis.
Features
- New – Bigger time stamp in upper right corner
- New – Standby button
- New – Revised SpO2 pitch tone is still audible when SpO2 rate decreases.
- New – Volume and Sounds are adjustable and now can be turned off completely
- New – Option to print Numerical Data Only
- New – Color of the waveforms can be changed
- EtCO2 Option allows compliance with ASA
- Available with optional dual invasive blood pressures, EtCO2, a battery and a built-in recorder.
- 12.1″ color TFT display with a touch-screen interface
- Multi-lead ECG monitoring
- Built-in battery (optional)
- NIBP list display
- OxyCRG dynamic view display
- Built-in thermal recorder
- Networkable with a central monitoring system via a wireless or wired network
- S-T Segment and 72 hours trend data analysis
- Compact design convenient for mobile monitoring
- Drug dose calculations
- Multi-lingual settings
SPECIFICATIONS
Physical Weight & Dimensions
- Weight: 5kg
- Height: ~11 inches
- Depth: ~6-1/5 inches
- Width: ~12 3/5 inches
Application
- Neonatal, pediatric and adult patients
Operation Environment
- Power Source – External AC power or internal battery
- AC Power – 100-240VAC, 50/60Hz, <150VA
- Battery – Rechargeable Lead-Acid; Operating time under the normal condition (1 hour); Operating time after the first alarm of low battery (10 min.)
Temperature
- Working – 5°-40° C
- Storage – -20° C
Performance Specifications
- Display – 12.1″ color TFT, Touch Screen
- Resolution – 800 × 600 pixels
- Trace – 4, 6 or 8 waveforms ECG (I, II, III, aVR, aVL, aVF, V1-V6), PLETH, RESP, IBPx2, ETCO2
- Indicator – Alarm indicator; Power indicator QRS beep and alarm sound
- Trend Time – 72 hours
- Recorder – Built-in, thermal array, 2 channels
- Record – Width 48mm
- Recorder – Paper 50mm
- Record – Speed 25mm/s, 50mm/s
ECG
- Input – 3-lead ECG cable and standard AAMI line for connection
- Input – (Optional) 5-lead ECG cable and standard AAMI line for connection
- Lead Choice – I, II, III, aVR, aVF, aVL, V
- Gain Choice – x0.5, x1.0, x2.0
- Frequency Characteristic – 0.05-40 Hz
- ECG Waveforms – 7 channels
- Penetration Voltage – 4000VAC 50/60Hz
- Sweep Speed – 12.5, 25 and 50 mm/s
- HR Display Range – 30-300bpm
- Accuracy – ±1% or ±1bpm, whichever is greater
- Alarm Limit Range Setting – Upper limit 80-400bpm; Lower limit 20-150bpm
RESP
- Measure Method – RA-LL impedance
- Range – 0-120 rpm
- Accuracy – + 3 rpm
- Alarm Upper-lower Limit Setting – Upper limit 6-120 rpm, Lower limit 3-120 rpm
- Sweep Speed – 12.5 and 25 mm/s
NIBP (Suntech)
- Measuring Technology – Automatic oscillating measurement
- Cuff Inflating – <30s-0-300 mmHg, standard adult cuff
- Measuring Period – AVE<40s
- Mode – Manual, Auto, STAT
- Measuring Interval in AUTO Mode – 2 min-4 hrs
- Pulse Rate Range – 30-250 (bpm)
- Measuring Range – Adult/Pediatric Mode, SYS 40-250 (mmHg), DIA 15-200 (mmHg); Neonatal Mode, SYS 40-135 (mmHg), DIA 15-100 (mmHg)
- Resolution – 1mmHg
- Accuracy – Pressure, Maximum Mean error ±5mmHg, Max. Standard deviation 8mmHg
- Overpressure Protection – Adult Mode 280(mmHg), Neonatal Mode 150 (mmHg)
- Alarm Limit Setting – SYS 50-240 mmHg, DIA 15-180 mmHg
TEMP
- Range – 25°-50° C
- Accuracy – + 0.2 °C 25.0°-34.9° C; + 0.1° C 35.0°-39.9° C; + 0.2° C 40.0°-44.9° C; 0.3° C 45.0°-50.0° C
- Display Resolution – 0.1° C
- Alarm Upper-lower Limit Setting – Upper limit 0-50° C, Lower limit 0-50° C
- Channel – 2 channels
- Alarm Limit Setting – 10-50° C
Masimo SET SpO2
- SpO2 Accuracy (non-motion):
- Adult Pediatric: 70~100 %: ±2 %, 0~69%: unspecified
- Neonate: 70~100 %: ± 3 %,0~69 %: unspecified
- SpO2 Accuracy (motion):
- Adult Pediatric: 70~100 %: ±3 %, 0~69%: unspecified
- Neonate: 70~100 %: ± 3 %, 0~69 % : unspecified
- Low Perfusion Performance: >0.02 % Pulse Amplitude and % Transmission >5
- SpO2: ±2 %
- PR: ±3 bpm
- Modes:
- Averaging mode: 2,4,8,10,12,14 and 16 s
- Sensitivity: Normal, APOD and Maximum
- PR Accuracy (non-motion):
- Neonate: 70~100 %: ± 3 %, 0~69 % : unspecified
- Adult Pediatric Neonate: 25~240 bpm: ±3 bpm
- PR Accuracy (motion):
- Adult Pediatric Neonate: 25~240 bpm: ±5 bpm
- Measuring Range:
- SpO2: 1~100 %
- PR: 25~240 bpm
- Perfusion: 0.02~20 %
IBP (Option)
- Measurement Range – -10-300mmHg
- Channel – 2 channels
- Pressure Transducer Sensitivity, 5μV/V/mmHg
- Impedance Range – 300-3000Ω
- Transducer Sites – ART, PA, CVP, RAP, LAP, ICP
- Resolution – 1mmHg
- Accuracy – ±1mmHg or ±2%, whichever is greater
- Alarm range – -10-300mmHg
EtCO2 (Option)
- EtCO2 Option allows compliance with ASA Standards for Basic Anesthetic Monitoring (Effective July 1, 2011)
- Mode of Sampling – Sidestream
- Principle of Operation -Non-dispersive infrared (NDIR) single beam optics, dual wavelength, no moving parts
- CO2 Measurement Range – 0 to 150 mmHg (0 to 19.7%, 0 to 20 kPa)
- CO2 Calculation Method – BTPS (Body Temperature Pressure Saturated)
- CO2 Resolution – 0.1mmHg(0-69mmHg); 0.25mmHg(70-150mmHg)
- CO2 Accuracy – 0-40 mmHg ± 2 mmHg; 41-70 mmHg ± 5% of reading; 71-100 mmHg ± 8% of reading; 101-150 mmHg ± 10% of reading; Above 80 breath per minute ± 12% of reading
- Sampling Rate – 100Hz
- Respiration Rate – 2-150 bpm
- Respiration Rate Accuracy – ±1 breath
- Response Time – <3 seconds – includes transport time and rise time
- Inspired CO2 Measurement Range – 3-50 mmHg